INFORMATION
■
By default, end of infusion pre-alarm is activated and will be triggered when
remaining time is 5 min. before end of infusion.
■
Time before end of infusion setting range is 1 ~ 59 min. (preset is 5 min.).
■
The remaining volume before end of infusion can be set from 1 mL to 999
mL (preset is 10 mL).
6.6.3 Nurse settings lock
This function enables to lock the pump to avoid misuse. The nurse settings lock is deactivated
by default.
When nurse settings lock is activated, the infusion parameters (flow rate, volume, duration,
etc.) cannot be changed.
1.
Under advanced settings menu (see
on page 60 to access),
select nurse settings lock menu
and press
(
) key to enter.
2.
Press
key to activate (or press
to deactivate) the nurse settings lock.
65