Actions
Alarm condition
Pumping
status
Symbol
■
Connect the pump to the
mains via the pump holder.
Minimum battery voltage
is not available. Appears
at least 30 min. before the
empty battery alarm.
Pumping
continues
Empty battery
pre-alarm
NOTE:
The
battery icon
becomes
Yellow
.
The number of bars
may vary,
depending on the
battery
performance.
■
Proceed to next step or switch
pump off.
Pump is switched on but
not operating for 2
minutes (3 beeps).
Pumping stops
Action reminder
Technical alarms
Actions
Alarm condition
Pumping status
Symbol
High priority - Red LEDs are flashing and alarm sound (sequences of 10 beeps(1 beep repeats
for power supply failure and sequences of 3 beeps for speaker failure))
84