3.
Confirm that the administration set is correctly installed in the pump.
4.
Open the roller clamp (for stationary set only).
5.
Connect the administration set to the patient's IV access device.
6.
Check the infusion settings prior to start the infusion.
7.
Press
to start infusion.
When the infusion starts, the pump displays the monitoring screen. To check the programmed
values without stop infusion, press
to enter the information screen. The information
screen will last 5 seconds then return to monitoring screen.
Information screen (last 5 s)
Monitoring screen
Programming screen
Infusion must be stopped to modify the current infusion parameters (more details, see
on page 36).
INFORMATION
The infusion parameters cannot be programmed on the information screen and
monitoring screen.
5.8 Keypad lock
Keypad lock prevents from unintentional change of pump settings during infusion.
INFORMATION
Keypad lock cannot be activated under following condition:
■
Priming is running.
■
Technical alarm occurs.
■
Empty battery alarm occurs.
34