4.1.3.
Cleaning the measurement chamber
First of all switch off the device and lift the measurement chamber (as described in chapter
4.1.2). There is a gap between the measurement chamber and the ground plate. Use this
gap and a thin brush or vacuum cleaner to clean the chamber.
Warning:
Be careful and d
on’t damage the sample arm inside the chamber.
The arm can be moved by hand using the motor wheel in the bottom area of the
measurement system (as shown in Figure 27).
Figure 27:
manual moving of the sample arm
The sample arm can be moved a quarter turn to clean the other parts of the chamber.
Hand wheel to move the
sample arm