application
. After clicking the
connection
- button
lexstudio
should connect to the
measurement system. If not, check the connection cables, recheck the IP address and the
network- settings of the operating system and make sure that the lexsyg measurement
system is powered and running. The IP-address of the connected PC has to be static. If you
forgot the correct IP address for your device, please call the support hotline of Freiberg
Instruments.
Figure 21: options menu
3.6.2.
Editing cooling power and evacuation time
In some particular cases it is necessary to change the power of the cooling air flow or the
duration of evacuation. This can be achieved by open the options window by clicking
Options
->
software options…
in the
menu bar (Figure 21). In the lower part of the window
the parameters can be changed.
3.6.3.
Breaking a measurement
Aborting a running measurement can be achieved by clicking the measurement button while
a running measurement. The software will ask if it is correct to abort and how to handle the
actual sample disc. When not moving the sample back to the sample wheel, keep in mind
that the disc must be removed manually.