User Manual 6007 2981 / Rev. 10
5:1
5
Maintenance
5.1
EyeFoss Maintenance
5.1.1
General
For the instrument to work properly a daily, weekly and annual service should be
made.
Daily and weekly maintenance are done by the operator of the instrument.
Annual maintenance is performed by a qualified service engineer.
Note: Do not use compressed air inside the instrument.
Daily:
•
A selftest should be done every morning, the simple way to do that is to set it for
automatic selftest.
•
“ZERO” should be done at least once a day.
Weekly:
•
Run the
Clean
procedure. It can be found in the Reports view under
Maintenance
.
Fig. 5:1
Simply follow the instructions on the screen.
5.1.2
Cleaning/replacement of Conveyor Belt
Consider the belt to be a wear part. The lasting time of the belt depends on the amount
of analysed samples and their cleanness.
When cleaning/replacing the conveyor belt, follow the steps below: