User Manual 6007 2981 / Rev. 10
4:19
14. Change the icon if required.
Fig. 4:44
15. Click
Save
button and the Quick job is finished.
This should now be run once a day to keep the instrument calibration updated.
16. Run a zero setting now.
4.7
Running a Self-test
A self-test should be made daily as a verification of the instruments performance and
quality.
It can be done either manually or automatically (recommended).
Note: If the automatic self-test is chosen, only errors in the self-
test will be displayed, not warnings. Warnings will only be
disaplyed if the self-test is completed manually.
Below is described how to do it manually.
1.
Go to the
Diagnostic
view in Foss Integrator and right-click on the instrument
icon (“Prediction PC Active”) in the lower left corner.
2.
Click on
Self Test
....
Fig. 4:45