Troubleshooting |
31
Error
Cause
Solution
n
Temperature fault occurred.
The Form Cure L thermistor
reported an error.
•
Disconnect and reconnect the power.
•
Disconnect the power. Remove
the Form Cure L top shell and
check that the thermistor wiring is
secure. Contact
or a
for
more information.
The LED brightness
encountered an error.
The Form Cure L
encountered an error while
setting the brightness of
the LEDs.
•
Disconnect and reconnect the power.
•
Disconnect the power. Unscrew
each of the three LED modules
(accessible from within the post-cure
chamber) and check that their wiring
is secure. Contact
or a
for
more information.
The set limit failed.
The Form Cure L
encountered an error
while setting the current
temperature limit.
•
Disconnect and reconnect the power.
•
Disconnect the power. Unscrew
each of the three LED modules
(accessible from within the post-cure
chamber) and check that their wiring
is secure. Contact
or a
for
more information.
The set duty cycle failed.
The Form Cure L
encountered an error while
setting the brightness of
the LEDs.
•
Disconnect and reconnect the power.
•
Disconnect the power. Unscrew
each of the three LED modules
(accessible from within the post-cure
chamber) and check that their wiring
is secure. Contact
or a
for
more information.
The door is open.
Please close.
The door is open while the
Form Cure L is trying to begin
a cycle.
Close the Form Cure L door.
The device failed to reach
the required temperature.
The Form Cure L did not
reach its target temperature
within 30 minutes.
Disconnect the power. Remove the
Form Cure L top shell and check that
the thermistor and heater module wiring
is secure. Contact
more information.
An error occurred with the
device’s turntable.
The turntable was unable
to move properly due to an
issue with the motor driver or
the motor command.
Disconnect and reconnect the power.
An unknown error occurred
Other/unspecified cause
Disconnect and reconnect the power.