EN-13
Fever Indicator
When the measurement is equal or over 38°C (100.4°F),
red backlight is shown together with the result. The red
backlight will last for 3 seconds and turn off.
Troubleshooting
The table below shows problems you may encounter.
All error messages below would be shown together with
orange backlight. Please follow “what to do” to resolve
problems. If the problem still exists, please call your
local dealer for help.
Message
What it means
What to do
Room temperature is
below 10°C.
Put the thermometer under
operating temperature range
of 10°C to 40°C (50°F to
104°F).
Room temperature is
above 40°C.
Put the thermometer under
operating temperature range
of 10°C to 40°C (50°F to
104°F).
You don’t use probe
cover while measuring
ear temperature.
Please place probe cover into
probe again.
Problem with the
thermometer.
Review the instructions and
re-start the measurement
procedure. If the above steps
do not work.
Please contact the dealer.
Battery is low and “
”
appears on LCD.
Please replace batteries as
soon as possible.
Appears when the
batteries can’t provide
enough power for a test.
Please replace new batteries.
Temperature
measurement
falls outside the
displayed temperature
range: (ear temperature
range from 32°C to
43°C. skin/surface scan
temperature range from
22°C to 44°C.)
Please follow this manual to
take a reading again.