26
MyCrown Mill
2.3.8
DC motors 'Touch'
Purpose of Test:
This test is used to check the two DC motors for proper functioning in the
Low-speed mode.
Procedure
The DC motors are tested by adjusting them to the touch speed (in the
relevant directions of rotation). Then the relevant data are determined.
DC motors 'Touch'
The possible motors can be selected either together or separately.
If no motor is selected, the
"Start"
button is not active. Default: Both
motors selected.
1
Measure of speed fluctuation
2
Pulse-width modulation
NOTICE
Different bearing friction
If the red fields in the service software drift from one side to the other
(e.g. from right to left) in the
"DC motors 'Touch'"
test after a spindle
motor is replaced, this may be attributable to different levels of bearing
friction in the two spindle motors. In this case, follow the procedure
described below:
➢ Start the
"Continuous load"
and let the spindle motors warm up for
approx. 15 min.
➢ Mill 1-2 restorations.
Summary of Contents for MyCrown Mill
Page 1: ...j j p j b Service Manual...
Page 15: ...15 Service Manual b Example Total test Total test...
Page 33: ...33 Service Manual b 3 1 Force measurement errors...
Page 34: ...34 MyCrown Mill 3 2 Stepping motor does not move...
Page 35: ...35 Service Manual b 3 3 Spindle motor does not lock...
Page 36: ...36 MyCrown Mill 3 4 Milling chamber door switch...
Page 37: ...37 Service Manual b 3 5 Unit switches off Continued on next page...
Page 38: ...38 MyCrown Mill Continued from previous page...
Page 40: ...40 MyCrown Mill Continued from previous page...
Page 41: ...41 Service Manual b 3 7 Insufficient air pressure...
Page 42: ...42 MyCrown Mill 3 8 Stepping motor cannot be controlled...
Page 43: ...43 Service Manual b 3 9 Faulty water pressure...
Page 44: ...44 MyCrown Mill Continued from previous page...
Page 45: ...45 Service Manual b 3 10 Unit cannot find starting position...
Page 46: ...46 MyCrown Mill 3 11 Unit cannot be turned on...
Page 47: ...47 Service Manual b 3 12 Touch errors...
Page 48: ...48 MyCrown Mill 3 13 Service life of the milling and grinding machine is too short...
Page 49: ...49 Service Manual b 3 14 Milling unit not addressable...
Page 50: ...50 MyCrown Mill 3 15 LAN communication problems Continued on next page...
Page 51: ...51 Service Manual b Continued from previous page...
Page 52: ...52 MyCrown Mill 3 16 WLAN connection occasionally interrupted...
Page 53: ...53 Service Manual b 3 17 WLAN communication problems...
Page 67: ...67 Service Manual b 4 5 Replacing stepping motors Overview of stepping motors...