2.3.2
Media supply
2.3.2.1
Media supply
Purpose of Test:
To check the media supply (air, water, fan) for proper functioning and test
the pressure switches and the function of the pump.
Procedure
The pumps and the fan are switched on by nominal select control
depending on selection. The motor currents and the condition of the
pressure switches are registered. The points of time when the pressure
detectors respond are measured. After approx. 5 seconds the pumps are
switched off and the time required until the pressure switch responds is
measured and evaluated.
Save
Save stores the measured values to the log file under the heading of
Media supply.
Media supply
Deviations of measurements
Deviations may occur:
● During the first test run (if any air is still in the water circuit)
● If the amount of water in the water tank is insufficient
● After filling the water tank.
Summary of Contents for MyCrown Mill
Page 1: ...j j p j b Service Manual...
Page 15: ...15 Service Manual b Example Total test Total test...
Page 33: ...33 Service Manual b 3 1 Force measurement errors...
Page 34: ...34 MyCrown Mill 3 2 Stepping motor does not move...
Page 35: ...35 Service Manual b 3 3 Spindle motor does not lock...
Page 36: ...36 MyCrown Mill 3 4 Milling chamber door switch...
Page 37: ...37 Service Manual b 3 5 Unit switches off Continued on next page...
Page 38: ...38 MyCrown Mill Continued from previous page...
Page 40: ...40 MyCrown Mill Continued from previous page...
Page 41: ...41 Service Manual b 3 7 Insufficient air pressure...
Page 42: ...42 MyCrown Mill 3 8 Stepping motor cannot be controlled...
Page 43: ...43 Service Manual b 3 9 Faulty water pressure...
Page 44: ...44 MyCrown Mill Continued from previous page...
Page 45: ...45 Service Manual b 3 10 Unit cannot find starting position...
Page 46: ...46 MyCrown Mill 3 11 Unit cannot be turned on...
Page 47: ...47 Service Manual b 3 12 Touch errors...
Page 48: ...48 MyCrown Mill 3 13 Service life of the milling and grinding machine is too short...
Page 49: ...49 Service Manual b 3 14 Milling unit not addressable...
Page 50: ...50 MyCrown Mill 3 15 LAN communication problems Continued on next page...
Page 51: ...51 Service Manual b Continued from previous page...
Page 52: ...52 MyCrown Mill 3 16 WLAN connection occasionally interrupted...
Page 53: ...53 Service Manual b 3 17 WLAN communication problems...
Page 67: ...67 Service Manual b 4 5 Replacing stepping motors Overview of stepping motors...