4.3 Adjusting nominal diameter
OPERATION
36
6.452
4.3.4 Fitting protective nozzles
Select nozzles according to Chapter. 4.2.1.
NOTE
!
Clean all parts before fitting them carefully!
Check for wear each time before using them!
Do not use worn nozzles!
Use only nozzle pairs with the same nominal diameter!
WARNING
!
Adjust the test heads to the nominal diameter of the protective nozzles
before inserting the nozzle!
Inserting the nozzle into a test head arrangement that is too narrow can
cause damage to the test heads and/or nozzles.
Entry nozzle 5 - 67 mm
Fit the entry nozzle (7) into the inserted adapter (6) by turning it.
Fit the clamping ring entry (8).
Tighten three Allen screws M 6 x 22 DIN 912 with 6 Nm (A).
Fig. 4.7 Fitting nozzles entry side, legend see Chapter 2.7