Operating Manual EMGZ309._.EIP
28
8.3
Description of Parameters
Display Filter
LCD:
Dis.Filt
Use:
The display filter (lowpass) steadies the value shown in the display
when there is a high fluctuation of the force values. The parameter
Dis.Filt
stores the cut off frequency of the filter. The lower the cut off
frequency, the more sluggish the display will be. The display filter does
not influence the output signal filter described in the
parameter
Out.
Filt
(see below).
Unit
Parameter Range
Selection
Default
Min
Max
Hz
0.1
10.0
-
1.0
Filter Amplifier Output
LCD:
Out.Filt
Use:
The EMGZ309 contains a lowpass filter to prevent noise overlaying to
the output signal. This parameter stores the cut off frequency of the
filter. The lower the cut off frequency, the more sluggish the output
signal will be. The filter stabilizes the output signal in case of highly
fluctuating force values. This filter is independent from the display filter
(see above).
Unit
Parameter Range
Selection
Default
Min
Max
Hz
0.1
200.0
-
10.0
Unit of Measure
LCD:
Unit
Use:
This parameter determines the unit system used
.
Note:
By changing the units to lb (pounds) the whole unit system will
change from metric to imperial units.
If the Device Mode [VoltGauge] is set, this parameter is deactivated
Unit
Parameter Range
Selection
Default
Min
Max
-
-
-
N
kN
lb
g
kg
N