Operating Manual CMGZ432
27
10.2 Command list
command answer
purpose
DAKT<CR>
XXXXXX<CR>
read actual diameter
DIFF<CR>
DIFY<CR> / DIFN<CR>
read alarm control error
DIFR<CR>
DIFRXXXX.X<CR>
actual control error in %
ERR?<CR>
XX<CR>
read actual erros
Pos.1...2 : Err1...Err2
Value of Pos. = 0 : No Err;
Value of Pos. = 1 : Err active
FREI<CR>
PACC<CR> / FAIL<CR>
enable controller
IDNT<CR>
CMGZ432 V2.05 0899
10 characters type, fix
< Typ > <Version> <S >
10 characters version, fix
4 characters serial number, fix
INRS<CR>
PACC<CR> / FAIL<CR>
initialize interface (for ex. after
loading of new interface
parameters)
LOCK<CR>
PACC<CR> / FAIL<CR>
disable conbtroller
REMR<CR>
PACC<CR> / FAIL<CR>
turn off remote mode
(enabling of the keys on
the operating panel)
REMS<CR>
PACC<CR> / FAIL<CR>
turn on remote mode
(disabling of the keys on
the operating panel)
SOLLXXXXXXX<CR>
PACC<CR> / FAIL<CR>
write new tension reference value
into RAM (The new reference will
be lost while power-off or
changing a parameter.)
SRMP<CR>
XXXXXX<CR>
read actual tension reference value
STEL<CR>
XXXX.X<CR>
read actual output value
SWRTXXXXXXX<CR> PACC<CR> / FAIL<CR>
write new tension reference value
into
EEPROM
STAR<CR> XXXXXXXXXXXXXXXXXX<CR> general status information
1-6 : feedback
7-11 : output value
12-16: reference after ramp
17 : error evaluation
18 : control error
VALS<CR>
XXXXXX<CR>
read feedback value
10.3 Read parameter
command answer
purpose
RP01<CR>
XXXXX<CR>
offset feedback
RP02<CR>
X.XXX<CR>
gain feedback
RP03<CR>
XXXX<CR>
sensor, nominal force
RP04<CR>
X<CR>
sensor, force unit
RP05<CR>
X.X<CR>
sensor, sensitivity
RP06<CR>
X<CR>
number of sensors
RP07<CR> XXX.X<CR>
lowpass,
feedback
RP08<CR> XX.X<CR>
lowpass,
instrument
RP09<CR> XX.X<CR>
lowpass,
display
RP10<CR>
XX.XX<CR>
scale of instrument
RP11<CR>
X<CR>
limit 1 min / max
RP12<CR>
XXXXX<CR>
limit value 1
RP13<CR>
X<CR>
limit 2 min / max
RP14<CR>
XXXXX<CR>
limit value 2