Expert Series Thermal Imager
Detailed Specifications
43
Detailed Specifications
Temperature Measurements
Temperature Range (not calibrated below -10
°
C)
TiX520 ............................................................ -20 °C to +850 °C (-4
°
F to +1562
°
F)
TiX560 ............................................................ -20 °C to +1200 °C (-4
°
F to +2192
°
F)
Accuracy
Range 1, 2 ......................................................
±
2
°
C or 2 % (whichever is greater) at 25
°
C ambient
Range 3 ..........................................................
±
4
°
C or 2 % (whichever is greater) at 25
°
C ambient
On-screen emissivity correction ......................... 1 % to 100 %
On-screen reflected background
temperature compensation ................................. yes
On-screen transmission correction ..................... 1 % to 100 %
Imaging Performance
Image Capture Frequency .................................. 9 Hz or 60 Hz refresh rate
Detector Type ..................................................... 320 X 240 Focal Plane Array, uncooled microbolometer
Thermal Sensitivity (NETD)
TiX520 ............................................................ 50 mK (40 mK with Filter Mode)
TiX560 ............................................................ 45 mK (30 mK with Filter Mode)
Total pixels ......................................................... 76,800
Infrared spectral band ......................................... 8
μ
m to 14
μ
m (long wave)
Visual (Visible Light) Camera
Type .................................................................... Industrial performance 5.0 megapixel
Minimum parallax alignment
with standard IR lens .......................................... ~60 cm (~24 in)