190 Series III
Service Manual
26
c. Press
2
(– READING …) and select with
1
(READINGS 2), and select on
B | V ac
.
d. Press
J
to select autoranging (AUTO in upper right LCD edge).
e. Use
j
and
k
to change the sensitivity range to select manual sensitivity ranging and
lock the Input B sensitivity range on
500 mV/div
.
f. Press
T
and use
1
to select Input B as trigger source.
3. Set the 5502A to source a sine wave at the first test point in
4. Observe the Input B reading and check that it is within the range shown under the
appropriate column of
5. Continue through the test points.
When you are finished, set the 5502A to Standby.
Input C AC Voltage Accuracy (HF) and Bandwidth Test
Note
The test steps for channel C are only for the 4-channel models.
To test the Input C high frequency automatic scope ac voltage measurement accuracy and
bandwidth:
1. Connect the Test Tool to the 5502A. See
.
2. Select the Test Tool setup:
). Press
S
,
2
(RECALL) and
select
SETUP
, press
E
, select the setup name, and press
E
to recall the
setup.
b. Press
P
, then press
2
(READING…) and select
READINGS 3 on C | V ac
.
c. Press
J
to select autoranging (
AUTO
in upper right LCD edge).
d. Use
j
and
k
to change the sensitivity range to select manual sensitivity ranging and
lock the Input C sensitivity range on
500 mV/div
.
e. Use
l
to move the Input C trace zero to the center grid line.
f. Press
T
and use
1
to select Input C as trigger source.
3. Set the 5502A to source a sine wave and to the first test point in
.
4. Observe the Input C reading and check that it is within the range shown under the
appropriate column of
5. Continue through the test points.
When you are finished, set the 5502A to Standby.