190 Series III
Service Manual
50
Table 14. Calibration Points
Cal Step
DUT Input Signal
5502A Setting
Input A LF-HF Gain Calibration Points (see
)
CL 0654
0.5 Vpp square wave, 1 kHz
SCOPE edge, 0.5 Vpp, 1 kHz
CL 0415
100 mVpp square wave, 1 kHz
SCOPE edge, 100 mVpp, 1 kHz
CL 0510
2.5 Vpp square wave, 1 kHz
SCOPE edge, 2.5 Vpp, 1 kHz
CL 0580
2.5 Vpp sine wave, 50 kHz
SCOPE levsine, 2.5 Vpp, 50 kHz
CL 0581
2.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 2.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0480
500 mVpp sine wave, 50 kHz
SCOPE levsine, 500 mVpp, 50 kHz
CL 0481
0.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 0.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0460
100 mVpp sine wave, 50 kHz
SCOPE levsine, 100 mVpp, 50 kHz
CL 0461
100 mVpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 100 mVpp,
501 MHz
381 MHz
151 MHz
111 MHz
Input B LF-HF Gain Calibration Points (see
)
CL 0674
0.5 Vpp square wave, 1 kHz
SCOPE edge, 0.5 Vpp, 1 kHz
CL 0435
100 mVpp square wave, 1 kHz
SCOPE edge, 100 mVpp, 1 kHz
CL 0530
2.5 Vpp square wave, 1 kHz
SCOPE edge, 2.5 Vpp, 1 kHz
CL 0582
2.5 Vpp sine wave, 50 kHz
SCOPE levsine, 2.5 Vpp, 50 kHz
CL 0583
2.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 2.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0482
500 mVpp sine wave, 50 kHz
SCOPE levsine, 500 mVpp, 50 kHz