ScopeMeter® Test Tool
Calibration Adjustment
41
Input A LF-HF Gain
To do the Input A LF-HF Gain calibration:
1. Connect Ch. A of the Test Tool to the 5502A. See
.
2. Press
2
to select the first calibration step in
,
Input A LF-HF Gain
.
The display must show step
CL 0654
(Pos A Fast).
3. Set the 5502A SCOPE output to source the signal required for the first calibration point.
4. Set the 5502A in operate (OPR) or standby (STBY) as indicated.
5. Press
3
to start the calibration.
Wait until the display shows calibration status
CL 509: READY
.
6. Press
2
to select the next calibration step, set the 5502A to the next calibration
point signal, and start the calibration.
Continue through all calibration points for
Input A LF-HF Gain
.
7. Wait until the display shows calibration status
CL 0461: READY
.
When you are finished, set the 5502A to Standby. Continue at the Input B LF-HF Gain section.
Input B LF-HF Gain
To do the Input B LF-HF Gain calibration:
1. Press
2
to select the first calibration step in
,
Input B LF-HF Gain
.
2. Connect Ch. B of the Test Tool to the 5502A. See
.
Figure 14. 5502A Scope Output to Test Tool Input B
3. Set the 5502A SCOPE output to source the signal required for the first calibration point for
Input B LF-HF Gain
(CL 0674, Pos B Fast).
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
A
D
C
B
USE 50
:
TERMINATION
A
B
METER
PM9091
2 CHAN. + METER
4 CHANNELS
50
:
(PM9585 or TRM50)
For 190-502 and -504
use only TRM50!
50
:
(PM9585 or TRM50)