ScopeMeter® Test Tool
Calibration Adjustment
53
Input ABCD Gain Calibration Points (see
)
CL 0799
5 mV, 50 Hz
Normal output
CL 0800
12.5 mV, 50 Hz
CL 0801
25 mV, 50 Hz
CL 0802
50 mV, 50 Hz
CL 0803
125 mV, 50 Hz
CL 0804
250 mV, 50 Hz
CL 0805
500 mV, 50 Hz
CL 0806
1.25 V, 50 Hz
CL 0807
2.5 V, 50 Hz
CL 0808
5 V, 50 Hz
CL 0809
12.5 V, 50 Hz
CL 0810
25 V, 50 Hz
CL 0811
50 V (set 5502A to OPR!), 50 Hz
CL 0812
125 V, 50 Hz
CL 0813
250 V, 50 Hz
DMM Meter Zero Calibration Points (2-channel models only)
CL 0890
no input, see
SCOPE Off, 0 V, 0 Hz, Operate
DMM Volt Gain Calibration Points (see
)
CL 0840
500 mV, 0 Hz
SCOPE Off, 500 mV, 0 Hz
CL 0849
2.5 V, 0 Hz
SCOPE Off, 2.5 V, 0 Hz
CL 0841
5 V, 0 Hz
SCOPE Off, 5 V, 0 Hz
CL 0842
50 V (set 5502A to OPR!), 0 Hz
SCOPE Off, 50 V, 0 Hz, Operate
CL 0843
500 V, 0 Hz
SCOPE Off, 500 V, 0 Hz
CL 0844
1000 V, 0 Hz
SCOPE Off, 1000 V, 0 Hz
Ohm Gain Calibration Points (see
)
CL 0910
100
Ω
SCOPE Off, 100
Ω
, Operate
CL 0911
1 k
Ω
CL 0912
10 k
Ω
CL 0913
100 k
Ω
CL 0914
1 M
Ω
CL 0915
10 M
Ω
CL 0916
100%: READY
Table 14. Calibration Points (cont.)
Cal Step
DUT Input Signal
5502A Setting