To test the Input C trigger sensitivity:
1. Connect the Test Tool to the 5502A. See
.
Note
The test steps for channel C are only for the 4-channel models.
Figure 8. 5502A Scope Output to Test Tool Input C
2. Reset the Test Tool to select the Test Tool setup.
3. Press
C
to turn Input C on.
4. Press
C
and use
l
to move the Input C trace zero to the center grid line.
5. Press
T
and use
1
to select Input C as trigger source.
6. Use
j
and
k
to change the sensitivity range to select manual sensitivity ranging and lock
the Input C sensitivity range on
2 V/div
.
7. Repeat step
to step
same as for the Input A trigger test.
When you are finished, set the 5502A to Standby.