![Fluke RUSKA 7252 User Manual Download Page 78](http://html1.mh-extra.com/html/fluke/ruska-7252/ruska-7252_user-manual_2310517078.webp)
RUSKA 7252
Users Manual
5-6
:DATA
:POINts?
return number of calibration constants
:VALue<n>?
returns label,value (C0,0.123)
:VALue<n> <number>
sets calibration constant
:CALibration
:VALue<n>?
return nominal calibration point
:POINts?
return number of calibration points
:ZERO <number>
Zero sensor
:PRESsure3?
RPT
Sensor
Same as PRESsure2 above
:PRESsure4?
A
PDCR
Same as PRESsure2 above
:PRESsure11?
B
Pressure*
Same as [PRESsure] above
:PRESsure12?
B
Case
Ref
Same as PRESsure2 above
:PRESsure14?
B
PDCR
Same as PRESsure2 above
:VACuum?
A
Vacuum
Same as PRESsure2 above
:VACuum11?
B
Vacuum
Same as PRESsure2 above
:TEMPerature?
A Gas Temperature
Same as PRESsure2 above
:TEMPerature2?
A Sensor Temperature
Same as PRESsure2 above
:TEMPerature11?
B Gas Temperature
Same as PRESsure2 above
:TEMPerature12?
B Sensor Temperature
Same as PRESsure2 above
DISP
:ENABle ON|OFF|1|0
turns front panel display on/off
:TEXT "<text>"
display message on front panel
[:PRESsure]
:BGRaph <number>
sets A bar graph maximum
:PRESsure11
:BGRaph <number>
sets B bar graph maximum
OUTPut
[:PRESsure]
:STATe ON|OFF|1|0
A off=MEASure, on=CONTrol
:MODE MEASure|CONTrol|VENT
A sets mode
:PRESsure11
:STATe ON|OFF|1|0
B off=MEASure, on=CONTrol
:MODE MEASure|CONTrol|VENT
B sets mode
PROGram
:CATalog?
returns list of defined programs
[SELected]
:DEFine <program block>
<step,press,toler,dwell,max>
:DELete
[:SELected]
deletes current program
:ALL
deletes
all
programs
Summary of Contents for RUSKA 7252
Page 3: ...7252 Change Language Hold key for 5 seconds enter mode ...
Page 4: ......
Page 10: ...RUSKA 7252 Users Manual vi ...
Page 12: ...RUSKA 7252 Users Manual viii ...
Page 20: ...RUSKA 7252 Users Manual 1 6 ...
Page 40: ...RUSKA 7252 Users Manual 3 6 ...
Page 72: ...RUSKA 7252 Users Manual 4 32 ...
Page 114: ...RUSKA 7252 Users Manual 6 22 ...
Page 118: ...RUSKA 7252 Users Manual 7 4 ...