5-1
Chapter 5
Remote Operation
Capabilities
The Calibrator can be operated remotely by a computer. Two interfaces are supported:
IEEE-488 and RS-232. Both interfaces support SCPI (Standard Commands for
Programmable Instruments). The IEEE-488 interface additionally supports emulation of a
RUSKA Single Channel Interface Panel (Models 6005-701 and 6005-761). The
IEEE-488 interface conforms to the following standards:
ANSI/IEEE Std 488.1-1987
IEEE Standard Digital Interface for Programmable
Instrumentation
ANSI/IEEE Std 488.2-1987
IEEE Standard Codes, Formats, Protocols, and Common
Commands
SCPI 1991.0
Standard Commands for Programmable Instruments
IEEE-488
The following identification codes define the interface capabilities of the Calibrator.
Identification codes are described in the IEEE-488 standard.
SH1
Source Handshake, Complete Capability
AH1
Acceptor Handshake, Complete Capability
T5
Talker
L3
Listener
SR1
Service Request, Complete Capability
RL1
Remote-Local,
Complete
Capability
PP0
Parallel Poll, No Capability
DC1
Device Clear, Complete Capability
DT0
Device Trigger, No Capability
C0
Controller, No Capability
The IEEE-488 interface is installed next to the processor board. The interface is identified
by the IEEE-488 standard connector on the back panel of the unit.
Note
Do
not change any jumpers or switch settings on the IEEE-488 interface
board. The IEEE-488 address is set by the MENU | SETUP | REMOTE
screen.
Summary of Contents for RUSKA 7252
Page 3: ...7252 Change Language Hold key for 5 seconds enter mode ...
Page 4: ......
Page 10: ...RUSKA 7252 Users Manual vi ...
Page 12: ...RUSKA 7252 Users Manual viii ...
Page 20: ...RUSKA 7252 Users Manual 1 6 ...
Page 40: ...RUSKA 7252 Users Manual 3 6 ...
Page 72: ...RUSKA 7252 Users Manual 4 32 ...
Page 114: ...RUSKA 7252 Users Manual 6 22 ...
Page 118: ...RUSKA 7252 Users Manual 7 4 ...