3-1
Chapter 3
Installation
Introduction
This section of the manual discusses initial installation for the RUSKA 7252 Calibrator.
Installing the Calibrator involves connecting the supply and test pressure tubing,
powering up the unit, and configuring the system through the front panel.
Unpacking the Calibrator
Carefully unpack all components, checking for obvious signs of damage. The shipment
contains the following items:
•
RUSKA 7252 or 7252i
Calibrator,
•
power cord,
•
user’s manual,
•
calibration report,
•
user-specified instrument options.
If necessary, report any shipping damage to the freight agency. Remove all shipping and
packing materials (including the shipping plugs) from all components. If possible, save
the packing materials for future shipping needs.
Finally, install the Calibrator in a location that meets the requirements listed in Table 3-1.
Note
The Calibrator should be handled in such a way as to minimize mechanical
shocks or vibration during installation or use. It should be mounted on a
rigid bench or in a sturdy 19" rack. Although the zeroing process will
compensate for a slightly unlevel mounting, the Calibrator should be
mounted to within 5° of level.