background image

M E D T E S T E R   5 0 0 0 C   F I E L D   U P G R A D E   I N S T A L L A T I O N   I N S T R U C T I O N S  

 

 

1

 

medTester 5000C Field 
Upgrade Installation 
Instructions 

These instructions tell you how to perform an instrument 
firmware upgrade to a medTester 5000C and how to install 
optional modules into a Fluke Biomedical medTester 5000C

Note 

If you ordered more than one module and/or upgrade, you have 
received more than one set of these instructions. But you can 
perform all installations at the same time with one set of instructions.  

Before You Begin 

In these instructions, 

upgrade

 means that you are bringing up to date the firmware 

in your medTester 5000. Upgrading involves replacing firmware integrated circuits 
(ICs). 

Enabling modules

 means that you are installing software from diskettes to your 

medTester in order to make medTester modules available for use. 

In some cases you need to install additional ICs that work with specific 
medTester 5000C modules. These are the modules for Expanded Memory, and 
medCheck. You have received all of the parts necessary to perform the installation 
yourself and to enable individual medTester 5000C modules.  

If you have already had Fluke Biomedical upgrade your medTester or install 
modules at the factory, you are receiving the module installation disks in the event 
that you need to reinstall the modules later. Please save these instructions with 
your installation disks. Your instrument is ready to use. 

 

Summary of Contents for Network Device MedTester 5000C

Page 1: ...mated Biomedical Equipment Test System Field Upgrade Installation Instructions PN 2245628 April 2005 2005 Fluke Corporation All rights reserved Printed in USA All product names are trademarks of their respective companies ...

Page 2: ......

Page 3: ...e firmware in your medTester 5000 Upgrading involves replacing firmware integrated circuits ICs Enabling modules means that you are installing software from diskettes to your medTester in order to make medTester modules available for use In some cases you need to install additional ICs that work with specific medTester 5000C modules These are the modules for Expanded Memory and medCheck You have r...

Page 4: ...oad them to your compatible CMMS or print them Note If you are installing a medTester 5000C firmware Upgrade and Modules 4 or 7 install your ICs before enabling the module s Module Identification Modules are identified by number according to the following table Module Number Module Name 2 RS 232 PRINTER 3 100 RECORDS 4 EXPANDED MEMORY 5 WAVES EXTENDED TEST 6 DATA TRANSFER 7 MEDCHECK 8 DEFIBRILLATO...

Page 5: ...ote If you are upgrading your medTester 5000C firmware and your medTester currently includes medCheck then your medCheck Module does not include the four 4 RAM ICs because they are already in your medTester However you may have received the installation disk in the event you need to reinstall it Required Equipment for Upgrade Modules To install a medTester 5000 firmware upgrade or to enable a medT...

Page 6: ...stalling 1 Any modules that include ICs to be installed inside the instrument including Firmware Upgrade Module 4 Expanded Memory Module 7 medCheck If you are not installing ICs skip to the section Enabling Modules below All ICs integrated circuits have a notch dot or dimple on one end This is the end with pin 1 All the ICs referenced here are oriented so that pin 1 is towards the rear of the inst...

Page 7: ...C Removal Hook the IC removal tool under each end of the IC and pull it out of the socket OR Insert a small flat bladed screwdriver under alternate ends and carefully pry them up until the IC comes out Plain Socket IC Installation Press the IC into the socket firmly Lever Handle IC Socket Removal Hold down the IC firmly with a finger and Pull the lever up Remove the IC by hand Lever Handle IC Sock...

Page 8: ... voltages present inside this instrument Always unplug the power cord before opening the case Never plug in the power cord while the case is open Disassembly Disassembly involves opening the medTester case in order to access ICs To disassemble 1 Peel off the Quality seal sticker that seals the case halves together Discard it 2 Remove the nine 9 screws and lockwashers holding the top case to the bo...

Page 9: ...rformed automatically B Calibration is not affected by this initialization so the instrument remains calibrated C The instrument should display a startup message then display MAIN MENU 1 D Test the RAMs From MAIN MENU 1 press F5 UTIL then RIGHT arrow then F2 DIAG then F4 MEMORY then F2 RAM The indicated RAM ICs should test GOOD All instruments U22 With Module 4 Expanded Memory U23 With Module 7 me...

Page 10: ... lowest numbered module you have Installing modules in numerical order will lead to a successful installation 1 Use the medTester 5000C COM1 port 2 Decide on a computer COM port COM1 through COM4 3 Decide on a baud rate 300 600 1200 2400 4800 9600 or 19200 4 To change the medTester baud rate from MAIN MENU 1 A Press F5 UTIL then F1 BAUD B Select the COM1 port C Press F4 BAUD until the desired baud...

Page 11: ...ter 9 The program will install enable the module in the medTester The computer should display MODULES SUCCESSFULLY LOADED Note If you have problems or get a different message you can read the README file by typing A README Enter for suggestions Modules are customized for each instrument and can only be installed on the instrument with that specific serial number Some modules need prerequisite modu...

Page 12: ...d Prompt by typing EXIT Enter at the DOS prompt Confirming Module Installation 1 Turn the medTester 5000 power off to reset the microprocessor 2 Turn the medTester 5000 power on Check that the startup message shows the correct instrument serial number and the correct firmware version number 3 Push F5 UTIL then right arrow then F3 MODULES At this point each press of F5 NEXT will sequentially scroll...