M E D T E S T E R 5 0 0 0 C F I E L D U P G R A D E I N S T A L L A T I O N I N S T R U C T I O N S
8
4.
Fold the case halves back together. Attach the top case to the bottom
case with the nine (9) screws and lockwashers. These are the stainless
steel screws.
Enabling Modules
Perform the steps described below for installing Modules 2 through 14. You need
to install each module from its separate diskette. Installation requires running a
program on an IBM-compatible personal computer.
Note
You
MUST
install the modules that you have received in numerical
order, starting with Module 2, or the lowest numbered module you
have. Installing modules in numerical order will lead to a successful
installation.
1.
Use the medTester 5000C COM1 port.
2.
Decide on a computer COM port—(COM1 through COM4).
3.
Decide on a baud rate—(300, 600, 1200, 2400, 4800, 9600, or 19200).
4.
To change the medTester baud rate from
MAIN MENU 1:
A.
Press
F5 UTIL
, then
F1 BAUD
;
B.
Select the COM1 port;
C.
Press
F4 BAUD
until the desired baud rate is displayed;
D.
Press
F5 STORE;
Press
Esc
twice to return to
MAIN MENU 1
.
Note
If Module 2 (RS-232/Printer) has not yet been installed, the
medTester baud rate is not selectable and can not be adjusted. It is
permanently set to 9600. After you install Module 2, you can change
the baud rate.
5.
Connect the computer COM port to the medTester COM1 port
(medTester to PC null modem serial cable).