M E D T E S T E R 5 0 0 0 C F I E L D U P G R A D E I N S T A L L A T I O N I N S T R U C T I O N S
9
6.
Run the computer in DOS mode (command prompt).
Note
If running Windows 95 or 98, with the mouse you can click and hold
the Start button, then select
Programs
and
MS-DOS Prompt
.
If running Windows 2000 or XP, with the mouse click the Start
button, select
Programs, Accessories,
and
Comnmand Prompt.
7.
Insert the diskette into the computer's 3.5" floppy disk drive.
8.
At the DOS prompt type this command:
A:\INSTALL port baud <Enter>
Where
A
is the floppy disk drive letter, A or B, and
port
is the computer COM port, 1 through 4, and
baud
is the selected baud rate: 300, 600, 1200, 2400, 4800, 9600, or
19200.
Here’s an example of this command:
A:\INSTALL 1 9600<Enter>
9.
The program will install (enable) the module in the medTester. The
computer should display
MODULES SUCCESSFULLY LOADED
.
Note: If you have problems or get a different message, you can read
the
README
file by typing
A:
README<Enter>
for suggestions.
Modules are customized for each instrument and can only be installed
on the instrument with that specific serial number. Some modules
need prerequisite modules to be installed first.
MODULE PREREQUISITE REQUIREMENTS
Module
Description
Prerequisite
2
RS232 / Printer
None
3
100 Record Storage
Module 2
4
Expanded Record Storage
Modules 2 & 3
5
Waveforms / Extended Testing
Modules 2 & 3
6
Data Transfer
Modules 2 & 3
7
MedCheck
Modules 2, 3, & 4
8
DEFIB Autosequences
Modules 2,3, & 4
9
IVPUMP Autosequences
Modules 2,3, & 4
10
Competitive CMMS Interface
Modules 2,3,4,6 &
7
11
ESU Autosequences
Modules 2,3, & 4
12
SPO2 Autosequences
Modules 2,3, & 4
13
PACER Autosequences
Modules 2,3, & 4
14
NIBP Autosequences
Modules 2,3, & 4
Summary of Contents for Network Device MedTester 5000C
Page 2: ......