Controller Operation
8.12 Calibration
Parameters
8
8-15
=
New linefeed setting
8.12 Calibration Parameters
The operator of the 9150 has access to the furnace calibration constants. These
values are set at the factory and must not be altered. The correct values are
important to the accuracy and proper and safe operation of the furnace. Access
to these parameters is available to the user only so that in the event the
controller memory fails, the user may restore these values to the factory settings.
The user should have a list of these constants and their settings with the manual.
Caution
DO NOT change the values of the furnace calibration constants
from the factory set values. The correct settings of these
parameters is important to the safety and proper operation of
the furnace.
The calibration parameters menu is indicated by,
Calibration
parameters
menu
Press “SET” five times to enter the menu. The calibration parameters menu
contains the parameters Hard Cutout, CT1, CE1, CT2, CE2, CT3, and CE3.
8.12.1 Hard Cutout
This parameter is the temperature above which the unit shuts down
automatically. The parameter is set at the factory to approximately 1260°C and
cannot be changed by the user.
8.12.2 CT1, CT2, and CT3
The calibration parameters CT1, CT2, and CT3 are the calibration temperatures.
8.12.3 CE1, CE2, and CE3
The calibration parameters CE1, CE2 and CE3 are the calibration errors
corresponding to the calibration temperatures.