Controller Operation
8.4
Ramp and Soak Program
8
8-7
Press “SET” to edit the program soak time.
Edit program point soak time
Current program point soak time
Use “UP”, “DOWN”, and “SET” to adjust the program soak time. This value can
be any integer from 0 to 14400. This time is the minutes the program set-point
maintains after the temperature of the furnace has settled and before proceeding
to the next set-point. Each digit flashes individually to indicate that it can be
adjusted.
Program point 4 soak time set for 200 minutes
Press “SET” to save the new soak-time value or “EXIT” to discard changes
Accept the program point soak time
The next value to edit is the program scan rate. This value is ignored if scan is
not enabled for the unit (See Section 8.3.1).
Program point 4 scan rate
Press “SET” to edit the program scan rate.
Edit the program point scan rate
Current program point 4 scan rate
Use “UP” and “DOWN” to adjust the program scan rate.
New program point 4 scan rate
Press “SET” to save the new scan rate value or “EXIT” to discard changes.
Accept the program point scan rate
After “SET” is pressed the controller advances to the next program point or, if
there are no more program points to edit, exits to the Program Function Menu.
Repeat the above steps to edit any program point.