Controller Operation
8.11Serial Interface Parameters
8
8-13
To change to manual reset mode press “UP” or “DOWN” and then “SET”.
Cutout set for manual reset
8.11 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. The parameters in the menu are: BAUD rate,
sample period, duplex mode, and linefeed.
8.11.1 BAUD Rate
The BAUD rate is the first parameter in the menu. The BAUD rate setting
determines the serial communications transmission rate.
The BAUD rate parameter is indicated by,
Serial BAUD rate parameter
Press “SET” to choose to set the BAUD rate. The current BAUD rate value is
then be displayed.
Current BAUD rate
The BAUD rate of the serial communications may be programmed to 300 600,
1200, 2400, 4800, or 9600 BAUD. 2400 BAUD is the default setting. Use “UP” or
“DOWN” to change the BAUD rate value.
New
BAUD
rate
Press “SET” to set the BAUD rate to the new value or “EXIT” to abort the
operation and skip to the next parameter in the menu.
8.11.2 Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature
measurements transmitted from the serial interface. If the sample rate is set to 5,
the instrument transmits the current measurement over the serial interface
approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. The sample period is indicated by,
Serial sample period parameter