Chapter 6: The Power Meter and Light Source
How to Monitor Power and Loss
95
GPU138.EPS
Figure 31. Connections to Monitor Power and Loss
To monitor loss
On the
POWER METER / LIGHT SOURCE
screen, tap
REF
to use the
present power level as a reference. The meters show the difference
between the reference power levels and the measured power levels.
To save the power and loss measurements
1
On the
POWER METER / LIGHT SOURCE
screen, tap
SAVE
.
2
On the
SAVE RESULT
screen, select
End 1
or
End 2
, make sure
the
Cable ID
and
End
name are correct, then tap
SAVE
. Saved
power meter results show these icons for
End 1
and
End 2
:
The tester saves the present, minimum, maximum, and reference
power and loss measurements for both wavelengths in one record.
Power at the
end of a link
Power from an
optical source
Optical network
interface card or
optical power source
Summary of Contents for CertiFiber Pro
Page 8: ...vi CertiFiber Pro Fiber Optical Loss Test Set Users Manual...
Page 42: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 32...
Page 50: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 40...
Page 92: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 82...
Page 98: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 88...
Page 110: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 100...
Page 130: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 120...
Page 142: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 132...
Page 144: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 134...
Page 150: ...CertiFiber Pro Fiber Optical Loss Test Set Users Manual 140...