Cardiac Functions
Blood Pressure Function
2
2-33
for BP channel 2, a delta change from 0 to 12;
for BP channel 3, a delta change from 0 to 5; and
for BP channel 4, a delta change from 0 to 3.
Action in the Menu-Control Mode
1.
Press the top-menu key labeled
BP
.
2.
Select
BP1, BP2, BP3,
or
BP4
to display the channel’s screen.
3.
Select
DYNA
to display the DYNAMIC WAVE screen for the selected channel.
4.
Select
ARTF
. On the LCD screen, OFF toggles to ON.
5.
Select
RUN
. On the LCD screen, RUN flashes to indicate the identified waveform is
running with the artifact applied.
6.
Press the
ESC
key to return to the selected BP channel screen. The LCD screen
identifies the waveform that is running and indicates if the artifact is active (ARTF
ON):
gje032.eps
7.
Press the
ESC
key again to return to the top menu BLOOD PRESSURE. The LCD
screen indicates that the selected waveform (in the following example, 120/80) on the
selected channel (2) is running with a respiration artifact (A prefixed to 120/80):
gje033.eps
Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-
artifact setting according to Table 2-26, and select
RUN
:
Table 2-26. Numeric Codes for Respiration-Artifact Settings
Respiration-Artifact Setting
Numeric Code
BP1 Respiration Artifact Off
349
BP1 Respiration Artifact On
350
BP2 Respiration Artifact Off
358
BP2 Respiration Artifact On
359
BP3 Respiration Artifact Off
367
Summary of Contents for Biomedical MPS450 Operators
Page 6: ......
Page 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Page 14: ...MPS450 Operators Manual viii...
Page 16: ......
Page 32: ...MPS450 Operators Manual 1 18...
Page 74: ...MPS450 Operators Manual 2 42...
Page 76: ...MPS450 Operators Manual 3 2...
Page 82: ...MPS450 Operators Manual 3 8...