MPS450
Operators Manual
4-12
Table 4-16. Numeric Codes for Performance Wave Amplitude Actions (cont.)
Numeric
Control Code
Action
Remote Control
Entry
281
Set performance amplitude to 0.50 mV.
PFA0.50
282
Set performance amplitude to 1.00 mV.
PFA1.00
283
Set performance amplitude to 1.50 mV.
PFA1.50
284
Set performance amplitude to 2.00 mV.
PFA2.00
285
Set performance amplitude to 2.50 mV.
PFA2.50
286
Set performance amplitude to 3.00 mV.
PFA3.00
287
Set performance amplitude to 3.50 mV.
PFA3.50
288
Set performance amplitude to 4.00 mV.
PFA4.00
289
Set performance amplitude to 4.50 mV.
PFA4.50
290
Set performance amplitude to 5.00 mV.
PFA5.00
291
Set performance amplitude to 5.50 mV.
PFA5.50
R-Wave Rate
Table 4-17 lists the numeric control codes for the R-wave rate functions.
Table 4-17. Numeric Codes for R-Wave Rate Actions
Numeric
Control Code
Action
Remote Control
Entry
312
Run R wave at 30 BPM.
RWR30
313
Run R wave at 60 BPM.
RWR60
314
Run R wave at 80 BPM.
RWR80
315
Run R wave at 120 BPM.
RWR120
316
Run R wave at 200 BPM.
RWR200
317
Run R wave at 250 BPM.
RWR250
R-Wave Width
Table 4-18 lists the numeric control codes for the R-wave width functions.
Table 4-18. Numeric Codes for R-Wave Width Actions
Numeric
Control Code
Action
Remote Control
Entry
318
Set R-wave width to 8 ms.
RWW8
319
Set R-wave width to 10 ms
RWW10
320
Set R-wave width to 12 ms
RWW12
321
Set R-wave width to 20 ms
RWW20
322
Set R-wave width to 30 ms
RWW30
Summary of Contents for Biomedical MPS450 Operators
Page 6: ......
Page 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Page 14: ...MPS450 Operators Manual viii...
Page 16: ......
Page 32: ...MPS450 Operators Manual 1 18...
Page 74: ...MPS450 Operators Manual 2 42...
Page 76: ...MPS450 Operators Manual 3 2...
Page 82: ...MPS450 Operators Manual 3 8...