Cardiac Functions
Cardiac Output
2
2-37
reads “0” or “24” on the CO monitor. (To get a reading, the trimpot may have to be
turned all the way in one direction.)
Note
Fluke Biomedical supplies adapters to connect to some nonstandard
catheters. (Please consult the cable price list.) For manufacturers that are
not listed, a “pigtail” may be ordered and wired to cut-off catheters.
Simulating a Cardiac-Output Test
The MPS450 simulates two separate injectate temperatures (an iced solution of 0
°
C / 32
°
F or a
room-temperature solution of 24
°
C / 75.2
°
F), each flowing at three different rates (2.5, 5.0, or
10 liters per minute).
Action in the Menu-Control Mode
1.
Press the top-menu key labeled CO to display the following LCD screen:
gje038.eps
2.
To set the injectate temperature to 0
°
C or 24
°
C, select
INJ
to toggle to the desired
setting.
3.
To adjust the flow rate, select
PREV
or
NEXT
to scroll to the desired setting.
4.
Select
RUN
. On the LCD screen, RUNNING flashes while the simulation is active.
To repeat a simulation, select
RUN
again.
5.
To terminate a cardiac-output simulation, select
END
.
Alternatively, in the Numeric-Control Mode, press the number keys for a combined
temperature/flow setting listed in Table 2-29, and select
RUN
:
Table 2-29. Numeric Codes for Temperature/Flow Settings
Temperature/Flow Setting
Numeric Code
CO OFF
090
CO 0 DEG 2.5 L/M
091
CO 0 DEG 5.0 L/M
092
CO 0 DEG 10.0 L/M
093
CO 24 DEG 2.5 L/M
094
CO 24 DEG 5.0 L/M
095
CO 24 DEG 10.0 L/M
096
Injectate Failure and Left-To-Right Shunt
Injectate failure occurs, for example, when an injector fails to release an injectate in a
continuous manner, a situation caused most often by the human factor of hesitation.
A physiological condition called a left-to-right shunt occurs when the blood detours from
Summary of Contents for Biomedical MPS450 Operators
Page 6: ......
Page 13: ...Contents continued vii List of Figures Figure Title Page 1 1 MPS450 Patient Simulator 1 8...
Page 14: ...MPS450 Operators Manual viii...
Page 16: ......
Page 32: ...MPS450 Operators Manual 1 18...
Page 74: ...MPS450 Operators Manual 2 42...
Page 76: ...MPS450 Operators Manual 3 2...
Page 82: ...MPS450 Operators Manual 3 8...