7 Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Set serial sampling setting to
n
seconds
sa[mple]=
n
sa=0
0 to 4000
Set serial duplex mode:
du[plex]=f[ull]/h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
Calibration Menu (WARNING – changing the following calibration values may change the accuracy of the instrument.)
Read C0 calibration parameter
*c0
*c0
c0: 9
c0: 0
Set C0 calibration parameter to
n
*c0=n
*c0=0
–999.9 to 999.9
Read CG calibration parameter
*cg
*cg
cg: 999.99
cg: 406.25
Set CG calibration parameter to
n
*cg=n
*cg=406.25
–999.9 to 999.9
These commands are only used for factory testing.
Read software cutout mode
*sco
*sco
sco: {ON or OFF}
sco: ON
Set software cutout mode:
*sco=ON/OF[F]
ON or OFF
Set software cutout mode on
*sco=ON
*sco=on
Set software cutout mode off
*sco=OF[F]
*sco=off
Miscellaneous (not on menus)
Read firmware version number
*ver[sion]
*ver
ver.9999,9.99
ver.9123,3.54
Read structure of all commands
h[elp]
h
list of commands
Legend:
[] Optional Command data
/ Alternate characters or data
{} Returns either information
n Numeric data supplied by user–may be entered in decimal or exponential notation
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Digital Interface Command Summary continued