2
Specifications and Environmental
Conditions
2.1
Specifications
Temperature Range
550°C to 1100°C (1022°F to 2012°F)
Accuracy
±3.0°C
Stability
±0.15°C
Uniformity
±0.05°C
Control Probe
10
Ω
HTPRT
Resolution
0.1°C/°F below 1000°C/°F
1°C/°F above 1000°C/°F
Readout
Switchable °C or °F
Controller
Digital controller with data retention
Fault Protection
High temperature cutout (Type R cutout thermocouple)
Sensor burnout and short protection
Cutout Accuracy
±10°C
Power
230 VAC (±10%), 50/60 Hz, 2500 W
Heater
2500 W
System Fuses
15 A 250 V fast acting
Exterior Dimension
838 mm H x 610 mm W x 406 mm D (33 x 24 x 16 in)
Weight
82 kg (180 lb.)
Safety
EN 601010-1, CAN/CSA C22.2 No. 601010.1-04
2.2
Environmental Conditions
Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
•
ambient temperature range: 5 - 50°C (41 - 122°F)
•
ambient relative humidity: maximum 80% for temperature <31°C, de-
creasing linearly to 50% at 40°C
•
pressure: 75kPa - 106kPa
•
mains voltage within ± 10% of nominal
7