25
ControllerOperation
SerialInterfaceParameters
parameters, serial interface parameters, and calibration parameters. The menus are
selected using the up and down arrows and then pressing “SET”.
6.9 Operating Parameters
The operating parameters menu is indicated by,
PAr
Operating parameters menu
The operating parameters menu contains the High Limit parameter.
6.10 High Limit
The High Limit parameter adjusts the upper set-point temperature. The factory default
and maximum are set to 375°C (707°F). The minimum setting is 125°C (257°F). For
safety, a user can adjust the High Limit down so the maximum temperature set-point is
restricted.
HL
Flashes
HL
and then displays the setting
375
Flashes the current value and then displays the value for
adjustment
375
Current High Limit setting
Press the
u
or
d
to adjust the setting.
300
New High Limit setting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
Accept the new High Limit setting
6.11 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SErL
Serial RS-232 interface parameters menu
Press “SET” to enter the menu. The serial interface parameters menu contains
parameters which determine the operation of the serial interface. The parameters in the
menu are — baud rate, sample period, duplex mode, and linefeed.
6.11.1 Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting determines the
serial communications transmission rate.
The baud rate parameter is indicated by,
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for 9100S
Page 8: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 16: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 18: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 26: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 38: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 42: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 44: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 48: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 50: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...