33
DigitalCommunicationInterface
InterfaceCommands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Secondary Menu
Read proportional band setting
pr[op-band]
pr
pb: 999.9
pb: 15.9
Set proportional band to
n
pr[op-band]=n
pr=8.83
0.1 to 30°C
0.2 to 54°F
Read heater power
(duty cycle)
po[wer]
po
po: 999.9
po: 6.5
Configuration Menu
Operating Parameters Menu
Read High Limit
hl[imit]
hl
hl: 9999.9
hl: 375.0
Set High Limit
hl[imit]=n
hl=300
125.0 to 375.0°C
257.0 to 707°F
Serial Interface Menu
Read serial sample setting
sa[mple]
sa
sa: 99999
sa: 1
Set serial sampling setting to
n
seconds
sa[mple]=n
sa=0
0 to 10,000
Set serial duplex mode:
du[plex]=f[ull]/
h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull]
du=f
Set serial duplex mode to half
du[plex]=h[alf]
du=h
Set serial linefeed mode:
lf[eed]=on/of[f]
ON or OFF
Set serial linefeed mode to on
lf[eed]=on
lf=on
Set serial linefeed mode to off
lf[eed]=of[f]
lf=of
Calibration Menu
Read R0 calibration parameter
r[0]
r
r0: 999.999
r0: 100.7
Set R0 calibration parameter to
n
r[0]=n
r=100.7
95.0 to 105.0
Read ALPHA calibration
parameter
al[pha]
al
al: 9.99999999
al: 0.003865
Set ALPHA calibration parameter
to
n
al[pha]=n
al=0.003865
0.002 to 0.006
Read DELTA calibration parameter de[lta]
de
de: 1.50
Set DELTA calibration parameter
de[lta]=n
de=1.37
de: 9.99999
0–3.0
Miscellaneous Other Commands
Read firmware version number
*ver[sion]
*ver
ver.9999x,9.99
ver.9100S,1.01
Read structure of all commands
h[elp]
h
list of commands
Read ALL operating parameters
all
all
list of parameters
Read setpoint resistance
*sr
*sr
999.999 ohms
110.023 ohms
Legend:
[] Optional command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed
by a carriage return and linefeed. Then the value is returned as indicated in the RETURNED
column.
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for 9100S
Page 8: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 16: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 18: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 26: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 38: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 42: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 44: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 48: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...
Page 50: ...1 888 475 5235 info Fluke Direct com Fluke Direct com...