116C
Users Manual
18
General Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18
°
C to 28
°
C, with
relative humidity at 0 % to 90 %.
Extended specifications are available at www.Fluke.com.
Maximum voltage between any
terminal and earth ground
.......................... 600 V
Surge Protection
......................................... 6 kV peak per IEC 61010-1600V CAT III,
Pollution Degree 2
Display
......................................................... Digital: 6,000 counts, updates 4/sec
...................................................................... Bar Graph: 33 segments, updates 32/sec
Temperature
................................................ Operating: -10 °C to + 50 °C
Storage: -40 °C to + 60 °C
Temperature Coefficient
............................. 0.1 x (specified accuracy)/
°
C (<18
°
C or >28
°
C)
Operating Altitude
....................................... 2,000 meters
Battery
.......................................................... 9 Volt Alkaline, NEDA 1604A / IEC 6LR61
Battery Life
.................................................. Alkaline: 400 hours typical, without backlight
Safety Compliances
.................................... Complies with ANSI/ISA 82.02.01 (61010-1) 2004,
CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003) and
IEC/EN 61010-1 2
nd
Edition for measurement Category III,
600 V, Pollution Degree 2, EMC EN61326-1
Certifications
............................................... UL,
P
, CSA, TÜV,
;
(
N10140), VDE