True-rms Multimeter
Rotary
Switch
Positions
7
Rotary Switch Positions
Switch Position
Measurement Function
x
Automatically selects ac or dc volts based on the sensed input with a low impedance
input.
e
Hz (button)
AC voltage from 0.06 to 600 V.
Frequency from 5 Hz to 50 kHz.
D
DC voltage from 0.001 V to 600 V.
l
AC voltage from 6.0 to 600 mV, dc-coupled.
DC voltage from 0.1 to 600 mV.
Ω
Ohms from 0.1
Ω
to 40 M
Ω
.
s
Continuity beeper turns on at
<
20
Ω
and turns off at
>
250
Ω
.
y
Temperature from -40
°
C to 400
°
C (-40
°
F to 752
°
F) with K-type thermocouple
R
Diode Test. Displays OL above 2.0 V.
S
Farads from 1 nF to 9999
μ
F.
$
DC current from 0.1 to 600
μ
A. AC current from 6.0 to 600
μ
A. DC-coupled.
Note: All ac functions and Auto-V LoZ are true-rms. AC voltage is ac-coupled. Auto-V LoZ, AC mV and AC
μ
A
are dc coupled.