B-1
Appendix B
Summary of Error Messages
Summary of Error Messages
Negative error numbers are from the Standard Commands for Programmable Instruments
(Version 1991.0).
Table B-1. Error Messages
Value
Description and Corrective Action
0 No
Error.
-103
Invalid Separator. Check punctuation in the SCPI command.
-104
Data Type. The type of parameter data is uncorrected. Check for numeric versus
string data.
-109
Missing Parameter. No valid parameter was found for the SCPI command.
-110
Command Header. The command name is not valid.
-113
Command Unknown. The command specified does not exist.
-114
Header Suffix. The numeric suffix for the command name is out of range.
-221
Settings Conflict. The command could not be executed due to the current state of the
DPC. Some commands cannot be executed while a program, self-test, or calibration
is in progress.
-222
Out of Range. The value is not within the valid range. For pressures check high and
low limits.
-281
Cannot create program. Program memory is full.
-282
Illegal Program Name. The name specified is not valid or does not exist.
-284
Program Currently Running. The command cannot be executed while a program is
running.
-285
Program Syntax Error. The syntax of the program definition is not correct.
-286
Program Runtime Error. An error occurred while running the program. Usually the
set-point is out of range.
Summary of Contents for RUSKA 7615
Page 8: ...RUSKA 7615 Users Manual vi...
Page 10: ...RUSKA 7615 Users Manual viii...
Page 16: ...RUSKA 7615 Users Manual 1 6...
Page 22: ...RUSKA 7615 Users Manual 2 6...
Page 40: ...RUSKA 7615 Users Manual 4 12...
Page 48: ...RUSKA 7615 Users Manual 5 8...
Page 62: ...RUSKA 7615 Users Manual A 4...