![Fluke Calibration RUSKA 7615 User Manual Download Page 55](http://html1.mh-extra.com/html/fluke-calibration/ruska-7615/ruska-7615_user-manual_2310769055.webp)
Maintenance
Calibration
6
6-7
Note
If the
actual pressure applied
is outside of the tolerance for the
requested high-point pressure, Error—222 Data out of range will occur.
Acknowledge this error by selecting
OK
, then re-enter the actual pressure,
repeating Step 3, number 1, if necessary
Step 4
1.
To begin Step 4, use the calibration standard to again
Apply
the mid-point pressure
of the range of the sensor being calibrated requested by the Calibrator. As pressure is
admitted into the Test Port, the
Measured pressure
on the Calibrator’s screen will
change accordingly.
2.
When the Measured pressure stabilizes, use the Calibrator’s numeric keypad and
OK
to enter the
actual pressure
reported by the calibration standard.
Do not enter the
Measured pressure reported by the Calibrator
. If necessary, use the
CLEAR
key to
correct a mistake in the edit field. If the
actual pressure applied
is acceptable, the
Calibration Complete
screen will appear.
Storing the Coefficients
Note
In addition to saving the calibration coefficients to the Calibrator’s memory,
the user is advised to separately record the calibration coefficients and
store this “backup” in a safe place.
Step 5
Calibration is complete. To exit the calibration procedure without storing the calibration
coefficients in memory, press
CANCEL
. To store the calibration coefficients in memory,
select
OK
.
Step 6
Press
PREV
to return to the main
screen.
Once the calibration procedure is complete, the user is advised to record several pressure
readings. If there are any significant errors at these points, then an error was probably
made in generating one of the calibration pressures, and the calibration procedure should
be repeated.
Editing the Calibration Coefficients
If the Calibrator’s memory is erased but the calibration coefficients are known, the user
can restore the coefficients to the Calibrator by following the directions below.
W
Caution
Never randomly adjust the calibration coefficients. Only
qualified personnel with valid backup data should be allowed to
edit the coefficients. If the backup coefficients are questionable,
perform the calibration procedure in its entirety.
1.
Verify that the Calibrator is in
Measure
mode (Chapter 4).
2.
To go to the
Calibration
screen, select
MENU/CAL
.
3.
To edit the calibration coefficients, press the recessed
CAL
button beneath the
vacuum fluorescent display. If the calibration access code is enabled, enter it at the
prompt. The
Calibration, Step 1
, screen will appear.
Summary of Contents for RUSKA 7615
Page 8: ...RUSKA 7615 Users Manual vi...
Page 10: ...RUSKA 7615 Users Manual viii...
Page 16: ...RUSKA 7615 Users Manual 1 6...
Page 22: ...RUSKA 7615 Users Manual 2 6...
Page 40: ...RUSKA 7615 Users Manual 4 12...
Page 48: ...RUSKA 7615 Users Manual 5 8...
Page 62: ...RUSKA 7615 Users Manual A 4...