HF197AEN_v0501_04
Page 26
Terminal 03-04; alarm output R2:
This output is the high alarm output.
Terminal 05-06; alarm output R1:
This output is the low alarm output.
Type OA:
An active 24V DC signal according to the functions R1 and R2 is available with this option.
Max. driving capacity 50mA@24V DC per output. (Requires power supply type PD / PF / PM).
Type OR:
:
A mechanical relay output according the functions R1 and R2 is available with this option.
Max. switch power 240V-0,5A per output. (Requires power supply type PF / PM).
+
INTERNAL
EXTERNAL
Active output - R1 and R2
6
5
R1
24V DC
max. 50mA
Common ground unit
R1 - GND
DEVICE
+
4
3
R2
24V DC
max. 50mA
Common ground unit
R2 - GND
DEVICE