1.
Safety information
1.1. Dangers of maloperation
The
CCM 01
underwent a safety inspection according to IEC 1010-1/ EN 61010
– 1, part 1.
Integrated hydraulic and electronic safety elements ensure safe operation if the device is
used as it was intended.
In case of maloperation or abuse, as well as in case of insensitivity for application
limits and safety regulations, the following threats can occur for:
Life or physical condition of the operator,
the
CCM 01
, as well as connected machines and systems connected,
the accuracy of measurements of the
CCM 01
,
the environment.
This manual does contain information and safety advice, which ensure risk free operations
and which help to keep the device in an ideal condition.
Therefore, it is necessary that everybody having to do with the operation and maintenance
of the unit strictly follows this instruction manual.
1.2. Intended applications
The
CCM 01
– Set
is an efficient and robust inline diagnostic measuring system for
determining the contamination classes according to ISO 4406:99 and NAS 1638.
It works very reliably and does fulfil all requirements of daily measurements.
The set is intended and tested to operate with all usual hydraulic and gear fluids as well as
synthetic esters.
Application limits:
Maximum acceptable operating pressure: 50 bar
/ 725 psi
Maximum oil temperature:
70 °C
/ 158 °F
Viscosity range:
10...400 mm²/s
/ 46...1854 SUS
Miniumum volume flow:
500 ml/ min
/ 0,132 gal/ min
Maximum acceptable volume flow:
50 l/min
/ 13 gal/min
Generally, the
CCM 01
– Set
has to be operated with 24 V DC (ripple: < 300 mV
pp
)!