22
*
marks the end of a data telegram
Test reading are displayed as follows:
$MEASURE;a.aa;b.bb;c.cc;d.dd;f.ff;dd;mm;yyyy;HH;MM;RR*
Means:
$MEASURE:
display the beginning of a new data field
a.aa:
particle per 1 ml
(ISO4406:
4 µm
(c)
)
or
particle per 1 ml
(NAS1638:
6,4 µm
(c)
= > 5 µm)
b.bb:
particle per 1 ml
(ISO4406:
6 µm
(c)
)
or
particle per 1 ml
(NAS1638:
14 µm
(c)
= > 15 µm)
c.cc:
particle per 1 ml
(ISO4406:
14 µm
(c)
)
or
particle per 1 ml
(NAS1638:
21 µm
(c)
= > 25 µm)
d.dd:
particle per 1 ml
(ISO4406:
21 µm
(c)
)
or
particle per 1 ml
(NAS1638:
37 µm
(c)
= > 50 µm)
f.ff:
volume flow through the measuring channel of the sensor
dd:
day (date)
mm:
month (date)
yyyy:
year (date)
HH:
hour (time)
MM:
minute (time)
RR:
report format (01: ISO, 02: NAS)
*:
marks the end of a data field
Example:
$M;371.08;29.44;2.92;0.74;50.00;02;01;2006;08;15;01*
3.3. Data transfer of the saved measurements
In the mode
„ Data transfer
“stored measurements can be transferred from the CCM 01 to
an external computer.
For this purpose, connect the CCM 01 using the serial interface RS232 with the external
Computer. (see chapter
3.3.1.
Export and evaluation of the stored data via
“CCM 01 data
manager Vers. 2
.0”
The CCM 01 DATA MANAGER software version 2.0 has been developed especially for
the CCM 01 display unit and is provided on the CD-ROM.
Connecting the CCM 01 to an external computer (see chapter 3.1) and
one-time
installation of the data manager software from the CD-ROM to the external computer is
necessary.
The data manager enables data transfer into ECXEL table.
3.3.2.
One-time program installation
Execute the CCM 01
– data manager program installation (
setup.exe
) from the
provided CD-ROM. The
setup.exe
is located in the following folder on the CD-ROM:
setup\ volume\ setup.exe.
Execute the installation as instructed and wait until the installation has been completely
finished.
In the Windows
– START – menu, the folder “CCM01 Data Manager” will be generated.
In this folder the data manager program
“CCM01 DataManager”
is saved.