User Manual
–
FloPatch FP120
Rev 1.0
CONFIDENTIAL
UNCONTROLLED UNLESS OTHERWISE SIGNED AND/OR STAMPED.
Page 32 / 33
fawf
(MHz)
4.03
4.03
4.03
4.03
Other
Information
prr
(Hz)
N/A
srr
(Hz)
N/A
npps
1
Ipa, α
at
zpii,α
(W/c
m
2
)
N/A
Ispta,α at
zpii,α or
zsii,α
(mW/
cm
2
)
15.78
Ispta at zpii
or zsii
(mW/
cm
2
)
33.07
pr at zpii
(MPa) 3.24E
-02
Operating
control
conditions
CW-mode
NOTE 1 Only one operating condition per index.
NOTE 2 Data should be entered for "at surface" and "below surface" both in the columns related to TIS
or TIB. NOTE 3 Information need not be provided regarding TIC for an TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
NOTE 4 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
NOTE 5 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the column
related to MI.
NOTE 6 Unshaded cells should have a numerical value. The equipment setting related to the index has
to be entered in the operating control section.
NOTE 7 The depths zpii and zpii,α apply to NON
-
SCANNING MODES, while the depths zsii and zsii,α
apply to
SCANNING MODES.
List of symbols
Symbol Term
MI
Mechanical index
TIS
Soft tissue thermal index
TIB
Bone thermal index
TIC
Cranial-bone thermal index
P
r,α
Attenuated peak-rarefactional acoustic pressure
f
awf
Acoustic working frequency
P
Output power
z
b
Depth for bone thermal index
z
s
Depth for soft-tissue thermal index
z
MI
Depth for mechanical index
prr