7.
Continue to hold the FloPatch securely
over the carotid and pull the liner tight to
apply tension over the skin. Once applied
tightly to the skin, pull sufficiently to rip
the liner off the end of the adhesive.
8.
Repeat Steps 6-7 for the liner on the
opposite side.
9.
After securing the FloPatch, the Doppler
wave trace can be viewed on the mobile
application. Perform any necessary
assessments and examine the mobile app
for information on patient physiology.
10.
Once the assessment is complete,
disconnect the FloPatch using the mobile
application or by physically pressing the
power button on the device for 2 seconds.
To disconnect using the mobile
application, click on the FloPatch icon on
the bottom left of the screen and press the
disconnect button on the screen that
appears.
The FloPatch may remain secured to the
patient to allow for reassessment.
11.
To perform subsequent assessments,
repeat Steps 1-2 to reconnect the device
to the mobile application.
Disconnect the device following the
assessment by repeating Step 10.
12.
After all assessments are completed, or
the device battery runs out, press the
power button on the device to turn it off if
it is still on. Remove the adhesive and the
device from the patient’s neck.
To remove
the adhesive, use an alcohol wipe/spray.
To remove the adhesive first spray the
alcohol on one edge of the adhesive, then
slowly start peeling the edge of the
adhesive as you spray between the
adhesive and the skin such that the
alcohol reaches between the adhesive and
the skin.
13.
Dispose of the device as required by local
regulations.
WARNING!
A single device is meant for single use on a single patient only. Attempting to re-
use a device can increase risk of cross contamination and cross infection.