Surge IEC 61000-4-5
±0.5 kV, ±1 kV
line to line &
±0.5 kV, ±1 kV, ±2 kV
line to ground
0%
U
T
(100% dip in
U
T
) for 0,5
cycle
Not Applicable
Not Applicable
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
0%
U
T
(100% dip in
U
T
) for 1
cycle(s)
70%
U
T
(30% dip in
U
T
)
for 30 cycles
0%
U
T
(100% dip in
U
T
) for 5 sec
Not Applicable
Not Applicable
Power Frequency
Magnetic Field
(50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.