EM6100 Handheld Portable Ultrasonic Flowmeter Ver.8.08
Flomotion
Systems,
Inc.
EM6100
36
J
Hardware Error
Hardware problem
Contact the factory
H
PoorSig Detected
(1)
Poor signal detected
(2)
Transducers installed improperly
(3) Too much fouling
(4) The pipe liners are too thick.
(5) Problem with transducers cords
(1) Relocate measuring place
(2) Clean the spot
(3) Check the cords
(4) Check the coupler
Q
Frequ OutputOver
The actual frequency for the
Frequency Output is out of the
range set by the user
Check the value entered at
M66,M67,M68 and M69, and
try to enter a larger value on
M69
F
System RAM Error
Date Time Error
CPU or IRQ Error
ROM Parity Error
(1) Temporary problems with RAM,
RTC
(2) Permanent problems with
hardware
(1) power on again
(2) contact factory
1
2
3
Adjusting Gain
Instrument is in the progress of
adjusting the gain for the signal, and
the number indicates the progressive
steps
K
Empty pipe
(1) No liquid inside pipe
(2) Setup error on M29
(1) Relocate where the pipe is
full of liquid
(2) Enter 0 on M29