EM6100 Handheld Portable Ultrasonic Flowmeter Ver.8.08
Flomotion
Systems,
Inc.
EM6100
27
Enter the lower flow rate value in window M68 and the higher flow rate value in window M69. Enter
the frequency range in window M67.
For example, assume that the flow rate varies in a range from 0 GPM to 3000 GPM, and an output
signal is at a maximum frequency of 1000Hz, the minimum of 200Hz is going to be required for other
instrumentation. The user should enter 0 in M68 and 3000 in M69, and enter 200 and 1000 in window
M67.
Please note that the user has to make the selection with OCT setups in window M78 by selecting the
13th option reading like ‘FO output’ to direct the frequency output to the OCT OUTPUT hardware
device.
§3.16 How to use the Totalizer Pulse Output
The totalizer output will produce a pulse output with every unit flow of the totalizer.
Refer §3.4 and §3.5 for the setups of the totalizer units and multiplier.
The totalizer pulse output can only be realized by mapping the pulse output to the OCT or BUZZER
hardware devices.
For example, assume that the POS totalizer pulse output is needed, and every pulse should represent
0.1cubic meter of liquid flow; the pulse output will be mapped to the internal Buzzer, so that with every
0.1 cubic meter of flow the BUZZER will beep for a while.
The following setups should be taken / performed:
(1) Select the unit Cubic Meter under window M32.
(2) Select the Multiplier as ‘2. X0.1’ under window M33.
(3) Select the output option ‘9. POS INT Pulse’ under window M77. (INT stands for totalized)
§3.17 How to produce an alarm signal
There are 2 types of hardware alarm signals that are available with this instrument. One is the Buzzer,
and the other is the OCT output.
Both for the Buzzer and OCT output the triggering sources of the event include the following:
(1) Alarms on when there is no receiving signal.
(2) Alarms on when there is poor signal received.
(3) Alarms on when the flow meter is not in normal measurement modes.
(4) Alarms on reverse flow.
(5) Alarms on the overflow of the Frequency Output.
(6) Alarms on when the flow is out of a designated range set by the user.
There are two out-of-normal-range alarms in this instrument. They are called #1 Alarm and #2 Alarm.
The flow range can be user-configurable through M73, M74, M75, M76.