OPERATION (continued)
USER CONFIGURATION MODE (Field Calibration Mode - continued)
Begin dispensing into a container of known
accurate volume. As soon as pulses are
detected by the software, the screen will
switch to display the volume being
dispensed on the top row and the volume
unit on the bottom row. (See Figure 31-3)
Figure 31-3
When the user is finished with the run, press and hold the
Total
button until the left
most digit begins to flash indicating it is in focus, then release the
Total
button.
Pressing the
Rate
button will increment the digit in focus. Pressing the
Total
button
will advance focus to the next digit.
Holding
Total
and pressing
Rate
will lock in the entered volume.
The software will then calculate the K-factor for the volume entered based on the
pulse count for the run and the volume entered. The average frequency and the K-
factor will be entered into the custom user K-factor table for run 1.
Exit Dispense Display Method
NOTE:
“FldCAL” will then be displayed on
the top row and “Exit” on the bottom row
(see Figure 32). The software will
temporarily store the sorted table of
frequencies and K-factors, then display the
“Save” screen. (See Figure 13)
Figure 32
NOTE:
Failure to save your entries on the “save” screen will cause the entries to be
discarded.